Volume 71, Issue 8 , Pages 399-405, August 2008
Sizing of Atrial Septal Defects by Intracardiac Echocardiography for Device Closures
Article Outline
Background
Transcatheter closure of a secundum atrial septal defect (ASD) has become an effective method for surgical treatment. In this study, we evaluated the feasibility and accuracy of intracardiac echocardiography (ICE) for sizing ASDs compared with conventional methods.
Methods
Between January 2004 and December 2006, 270 patients underwent transcatheter closure of secundum ASD by using septal occluders. For 142 patients, the procedure was guided by transesophageal echocardiography (TEE), and for 128 patients by ICE. We compared the maximal diameters of the ASDs obtained during angiocardiography, transthoracic echocardiography (TTE), ICE, and TEE with balloon-stretched sizes ascertained by using a sizing plate.
Results
ASD diameters measured with the sizing plate were significantly correlated with those measured with ICE (r = 0.963), TEE (r = 0.912), angiography (r = 0.88), and TTE (r = 0.85). The predicted stretched diameter of the ASDs, i.e. (nonstretched diameter measured with ICE ×1.07) +3.23 mm, agreed well with that measured by using a sizing plate (Ri = 0.974).
Conclusion
ASD diameters measured with ICE correlated with sizing-plate measurements better than those determined with TEE, angiography or TTE.
Key Words: atrial septal defect , intracardiac echocardiography , transcatheter closure
No full text is available. To read the body of this article, please view the PDF online.
References
- . Transcatheter closure of secundum atrial septal defects using the new selfcentering Amplatzer septal occluder: initial human experience . Cathet Cardiovasc Diagn . 1997;42:388–393
- . Intracardiac echocardiographic guidance during transcatheter device closure of atrial septal defect and patent foramen ovale . Mayo Clin Proc . 2004;79:24–34
- . Short and intermediate-term results of transcatheter closure of atrial septal defect with the Amplatzer septal occluder . Am Heart J . 2004;148:511–517
- . Transcatheter therapy of Lutembacher syndrome . J Chin Med Assoc . 2007;70:253–256
- . Transesophageal echocardiography guidance of transcatheter closure of atrial septal defect . Am J Cardiol . 1990;66:207–213
- . Twodimensional echocardiographic and color Doppler measurements of atrial septal defect, and comparison with the balloon stretched diameter . Am J Cardiol . 1993;72:1095–1097
- . Transesophageal echocardiographic results of catheter closure of atrial septal defect in children and adults using the Amplatzer device . Am J Cardiol . 2000;85:1387–1390
- . Intracardiac echocardiography: an ideal guiding tool for device closure of interatrial communications . Eur J Echocardiography . 2005;6:92–96
- . Comparison of costs of intracardiac echocardiography and transesophageal echocardiography in monitoring percutaneous device closure of atrial septal defect in children and adults . Am J Cardiol . 2004;94:690–692
- . Interatrial defect sizing by intracardiac and transesophageal echocardiography compared with fluoroscopic measurements in patients undergoing percutaneous transcatheter closure . Catheter Cardiovasc Interv . 2004;62:415–420
- . Echocardiographic guidance of transcatheter closure of atrial septal defects: is intracardiac echocardiography better than transesophageal echocardiography? . Pediatr Cardiol . 2005;26:135–139
- . Transcatheter closure of atrial septal defects and patent foramen ovale under intracardiac echocardiographic guidance: feasibility and comparison with transesophageal echocardiography . Cathet Cardiovasc Intervent . 2001;52:194–199
- . Intracardiac echocardiography guided device closure of atrial septal defects . J Am Coll Cardiol . 2003;41:285–292
- . Transcatheter closure of atrial septal defect and interatrial communications with a new self expanding nitinol double disc device (Amplatzer septal occluder): multicentre UK experience . Heart . 1999;82:300–306
- . Intracardiac echocardiography in congenital heart disease: are we ready to begin the fantastic voyage? . Pediatr Cardiol . 2002;23:286–291
- Sizing of atrial septal defects to predict successful closure with transcatheter CardioSEAL device . Tex Heart Inst J . 2001;28:177–182
- . Intracardiac ultrasound assessment of atrial septal defect: comparison with transthoracic echocardiographic, angiocardiographic, and balloonsizing measurements . Cardiovasc Intervent Radiol . 2001;24:84–89
- Transcatheter atrial septal defect closure: modified balloon sizing technique to avoid overstretching the defect and oversizing the Amplatzer device occluder . Cathet Cardivasc Intervent . 2005;65:390–396
- . Measurement of atrial septal defect size: a comparative study between three-dimensional transesophageal echocardiography and the standard balloon sizing methods . Pediatr Cardiol . 2000;21:465–469
- . Intracardiac echocardiography . Eur J Echocardiography . 2001;2:234–244
- . Use of intracardiac echocardiography to guide catheter closure of atrial communications . Echocardiography . 2003;20:781–787
- . Comparison of intracardiac echocardiography versus transesophageal echocardiography guidance for percutaneous transcatheter closure of atrial septal defect . Am J Cardiol . 2004;93:437–440
- . Transcatheter atrial septal defect closure assisted by intracardiac echocardiography: 3-year follow-up . J Interven Cardiol . 2004;17:95–98
- . Role of intracardiac echocardiographic guidance in transcatheter closure of atrial septal defects and patent foramen ovale using the Amplatzer device . J Interven Cardiol . 2003;16:51–62
- . Residual inferior atrial septal defect after surgical repair: closure under intracardiac echocardiographic guidance . J Invas Cardiol . 2001;13:810–813
- . Real-time threedimensional intracardiac echocardiography . Ultrasound Med Biol . 2001;27:1177–1183
PII: S1726-4901(08)70090-X
doi:10.1016/S1726-4901(08)70090-X
© 2008 Elsevier. Published by Elsevier Inc. All rights reserved.
Volume 71, Issue 8 , Pages 399-405, August 2008